Report No.:
Model:
TL1-A 1-11923
75 °F +/- 2º
MS
Calibration Data As Found
Calibration Data As Left
In
Tolerance
º F
º C
º F
º C
º F
º C
º F
º C
º F
º C
º F
º C
32
0
32.011
0.006
32.03
0.02
-0.02
-0.01
0.10
0.06 Yes
0.030
0.017
120
49
120.015
48.897
120.02
48.90
0.00
0.00
0.10
0.06 Yes
0.030
0.017
199
93
199.022
92.790
199.05
92.81
-0.03
-0.02
0.10
0.06 Yes
0.030
0.017
300
149
300.230
149.017
300.23
149.02
0.00
0.00
0.10
0.06 Yes
0.030
0.017
R0: 100.00
A: B: C:
Date:
Callendar Van Dusen Coefficients:
The results stated on this report relate only to the items specifically identified.
6/4/2013
This test report or calibration certificate shall not be reproduced except in full,
without written approval of the laboratory.
3.91400E-03
-6.03510E-07
Tested temperatures and corrections are as follows:
Tolerance
Correction
Measurement
Uncertainty
2013-06-04
Unit SN:
- 1-11923
New Unit or no "As Found" data available
This device has been adjusted to read as closely as possible to actual temperature.
Reading of TL1
Actual Test Temp.
Nominal Value
Calibration Date:
Ambient Temp:
Calibrated By:
This report is to certify that the instrument listed below
has been calibrated by ThermoProbe, Inc. to NIST
traceable criteria.
Report of Test
Traceability: This calibration is traceable to NIST through an unbroken chain of comparisons.
IsoTech calibrated the references shown above using transfer standards which in turn were
calibrated by their primary reference thermometer.
Uncertainty Statement: Uncertainties were computed using the concepts, methods and
techniques of the ISO Guide to the Expression of Uncertainty in Measurement (the GUM). The
calculated uncertainty is an expanded uncertainty (k=2). It does not consider errors due to
possible damage to the TL1 from shipping, temperature drift, or thermal hysteresis effect. To
maintain the accuracy of the TL1, users should take care to protect it during shipping, avoid using
it to measure temperatures significantly above the highest calibrated temperature, and have the
TL1 recalibrated annually.
Calibrator's Signature: ___________________________________________
Test Results Approved by: _______________________________________
Test Method: The calibration procedures used were ThermoProbe, Inc. Calibration Procedures based on ASTM E-
644-06. This probe was immersed in a constant temperature bath with a reference thermometer which determined
the actual test temperature. The readings were compared and correction factors for the probe were calculated. The
As Left readings reflect the TL1's readings after calibration
Report of Test
ThermoProbe
Bath Fluid
Reference Calibration Date
Next Calibration
Due
(-)20.0º C (-)4.0º F Fluke 7340 water/glycol TL1R-6068 3/27/2013 3/27/2014
0.0º C 32.0º F Fluke 7340 water/glycol TL2-0016 3/27/2013 3/27/2014
48.9º C 120.0º F PolyScience 8101 mineral oil TL2 - 0008 7/20/2012 7/20/2013
92.8º C 199.0º F Fluke 6330 mineral oil Fluke 1502A - ASP WSP500 3/27/2013 3/27/2014
149.0º C 300.2º F Fluke 6330 Silicon oil Fluke 1502A - ASP WSP500 7/20/2012 7/20/2013
290.0º C 554.0º F Fluke 9172 metrology well Fluke 1502A - ASP WSP500 7/20/2012 7/20/2013